Purite Integra 200E
Purite Integra 200E
For clinical laboratory analyser feed
Type II / Deionised (DI) water
The Purite Integra 200E water purification system provides high quality, high demand water for laboratories. The system uses the latest low energy reverse osmosis membranes and electro-deionisation technology to improve water quality and minimise operating costs.
- Completely self-contained, skid-mounted
- ASTM Type I, II or III water and BS EN ISO grade 1,2 or 3 water
- Make-up flow rate 200 l/hr
- Advanced monitoring and control
- Automated chemical cleaning program
Overview:
Supplied fully tested and ready for fast installation, the Purite Integra 200E water purification system provides high volume capability to help laboratories scale-up to meet increasing demand.
The system feature a range of advanced technologies, such as variable speed pumps and low-energy reverse osmosis membranes, to minimise operating costs. A range of options are available, including 350, 500 and 1000 litres water storage tanks.
Features:
- Self-contained, fully bunded, “plug and play” package designed to reduce installation and service times
- Efficiently delivers highly purified water; make-up flow rate of 200 l/hr
- Utilises reverse osmosis, electro-deionisation, ultra-violet irradiation and bacterial microfiltration
- Standby mode and variable speed pumps to minimise power consumption and running costs during periods of low demand
- LCD touch screen with password controlled menu access
- Automatic alarm notification system monitoring leakage and quality of water
- Compact, fully bunded, stainless steel skid package with anti-vibration mounts
- Category 5 backflow prevention to protect mains supply
- Automated chemical cleaning program
- Optional water storage available, 350, 500 and 1000 litres.
Tech Specs
| Unit specification | 200E – 350 | 200E – 500 | 200E – 1000 |
|---|---|---|---|
| Width (mm) | 1500 | 1850 | 2600 |
| Depth (mm) | 2020 | 2020 | 2020 |
| Height (mm) | 1020 | 1020 | 1020 |
| Maximum Shipping Weight (kg)* | 340 | 350 | 370 |
| Maximum Working Weight (kg)* | 767 | 917 | 1417 |
| Installation requirements | |
|---|---|
| Power | Single Phase, 110-230V, +/- 10%, 50 Hz |
| Feed Water | Softened |
| Maximum TDS (ppm) | < 1000ppm |
| Minimum inlet pressure – psi (bar) | 30 (2.1) |
| Maximum inlet pressure – psi (bar) | 90 (6.2) |
| Feedwater temperature | 10-25oC |
| Flowrate | 400 l/hr |
| Free chlorine | Must be dechlorinated |
| System specification | 200E – 350 | 200E – 500 | 200E – 1000 |
|---|---|---|---|
| Pure water storage | Up to 350 litres | Up to 500 litres | Up to 1000 litres |
| Display panel | LCD – Colour touch screen | ||
| Pre-treatment | 5μm pre-filter | ||
| Reverse osmosis | Low energy membranes | ||
| Deionisation | EDI module | ||
| Micro filtration | 0.2μm | ||
| UV lamp | Bactericidal 254nm | ||
| Purified water make-up flowrate @ 10°C | 200 l/hr | ||
| Purified water distribution | Up to 2m3/hr and a maximum of 90 psi (6.2 bar) | ||
| Carbon dioxide degassing | Hollow fibre membrane as standard | ||
| TOC reduction | Yes | ||
| Treated water specification | |
|---|---|
| Conductivity | Up to 15 MΩ-cm |
| Bacteria | < 1cfu/ml |
| Organics – TOC (ppb) | < 500ppb as C |
| Particles | < 0.2μm |
| Endotoxins | – |