Purite HPA
Purite HPA
For clinical laboratory analyser feed
Type II / Deionised (DI) water
The Purite HPA 30 provides a guaranteed and consistent flow of purified water to a quality better than 10MΩ-cm. The unit is compact, uses proven technology and is low maintenance, making it ideal for supplying all leading brands of clinical analysers.
- Make-up production rate of 30l/hour
- Optional remote display pod
- Energy saving intelligent stand-by mode
- Colour touch screen display with process graphics
- Wi-Fi enabled for remote monitoring and operation
Overview:
The Purite HPA 30 integrates reverse osmosis, deionisation and micro-filtration, plus a sophisticated control and colour LCD touch-screen display in a compact, easy to install and use unit.
An internal 20 litre storage tank is fitted as standard, with the option of larger external tanks. The HPA 30 is compatible with all major brands of clinical analysers.
Features:
- Guaranteed > 10MΩ.cm water quality
- Remote Display Pod (Optional)
- Energy saving intelligent stand-by mode
- Manual dispense from storage tank
- Colour touch screen display with process graphics
- Water quality parameters, MΩ.cm, °C, flowrate displayed
- Internal microfiltration
- Make-up production rate of 30 litres per hour (@10 °C)
- 20, 50 and 100 litre storage options
- Can be bench, under bench or wall mounted
- WiFi enabled for remote monitoring and operation
- 8mm pressurised outlet for direct connection to analyser.
Tech Specs
Unit specification | HPA 30 |
---|---|
Width (mm) | 440 |
Depth (mm) | 560 |
Height (mm) | 750 |
Maximum Shipping Weight (kg)* | 41 |
Maximum Working Weight (kg)* | 59 |
Installation requirements | HPA 30 |
---|---|
Power | Single Phase, 110-230V, +/- 10%, 50/60 Hz |
Feed Water | Potable |
Maximum TDS (ppm) | 1000 |
Minimum inlet pressure – psi (bar) | 30 (2.1) |
Maximum inlet pressure – psi (bar) | 90 (6.2) |
Make up outputs @ 10ºC (l/hr)* | 30 |
Make up outputs @ 25ºC (l/hr)* | 48 |
*Outputs based on a feed water pressure of 4 bar
System specification | HPA 30 |
---|---|
Pure water storage | 20 litre storage tank as standard (External 50 & 100 litre tanks available) |
Display panel | LCD – Colour touch screen |
Pre-treatment cartridge | Yes |
Reverse osmosis | Yes |
Deionisation cartridge | Yes |
Internal filtration | Microfiltration |
Recirculation pump | Yes |
Treated water specification | HPA 30 | Purified Water Storage Tank |
---|---|
Inorganics | > 10MΩ.cm |
pH* | Neutral |
Bacteria | < 1cfu/ml |
Organics – TOC (ppb) | < 20 |
Particles | 0.005µm |
Pressurised outlet (8mm) | Up to 2l/min @ 2.3-2.5 bar |
*pH of stored water may decrease due to absorption of free carbon dioxide